Description
Fields of Application
- chemical industry
- laboratory research
- quality control of technological processes in high-precision, science-intensive manufacturing.
Intended Use
- HygroTrace is created for the quantitative analysis of trace moisture in gases.
- The use of a coulometric sensor ensures a quick response of the analyzer to changes in gas parameters, thereby providing constant process monitoring and the ability to promptly adjust parameters in case of an emergency situation.
Signature Features
- Low Detection Limits
Measurement range for volumetric moisture content:- Range I: from 0.1 to 2,000 ppm
- Range II: from 1 to 2,000 ppm
- Range III: from 5 to 2,000 ppm
Measurement range for mass concentration of moisture: from 0.07 to 1500 mg/m³
- Wide Range of Dew Point Temperature Detection
from -90 to 20 ºC - High Energy Efficiency
Power consumption not exceeding 15 W - Eco-friendliness
Gas flow through the measurement chamber: from 0.05 to 0.1 L/min - High Accuracy
Absolute error limit in the range from 0.1 to 1 ppm: ±0.1
Permissible relative error limit in the range from 1 to 2,000 ppm: ±7.5%
Absolute error limit for dew point temperature measurement: ±1 ºC - Other Features
Maximum operating gas pressure up to 16 MPa (when using LNG Model-018)
Analog and digital data output channels