Description
The GPS provides an option for installing a reference instrument on a fold-out console. Gas quality analyzers installed on the GPS (such as hygrometers) can be used for measuring dew point under operating pressure, contract pressure or at a pressure of 2.7 MPa (the pressure at which hydrocarbon condensation temperature is optimally measurement).
Mounting points for the main analyzer and the reference instrument are fully adapted for the CONG-Prima and Hygrovision series dew point analyzers, respectively.
To expand the capabilities of the GPS, the supplied set can include various supplemental kits and systems: i.e. a pipeline connection kit, a high-pressure gas venting kit, measurement kit for both operating and reduced gas pressures, and the Model-001 additional supplemental cooling system.
Signature features
- Operation under various pressure conditions
The system works equally well at pipeline operating pressures (max 16.0 MPa), low pressures (8.0 to 0.1 MPa) and/or contract pressure (for hydrocarbon condensation temperature measurement) - Unique design
- Quick connections and fold-out console for installing a Hygrovision series reference instrument; Outlet for gas “flaring” for reference instrument
- High degree of gas purification
High degree of purification without distorting the gas sample is provided by means of an inertial-gravitational filter